Archive

1 posts

Showing results for: ICD

In vitro tests reveal sample radiofrequency identification readers inducing clinically significant electromagnetic interference to implantable pacemakers and implantable cardioverter-defibrillators.

Research RF Safe Research Library Jan 1, 2010

This in vitro study tested electromagnetic compatibility between passive RFID readers and 15 pacemakers plus 15 ICDs across 134 kHz, 13.56 MHz, and 915 MHz bands. EMI reactions were common with low-frequency RFID and rare with high-frequency RFID, while no reactions were observed with ultra-high-frequency or…

Page 1 / 1